EASY QC BLOOD GAS AND ELECTROLYTE QUALITY CONTROL
LEVEL 1 – REF 6303
LEVEL 2 – REF 6304
LEVEL 3 – REF 6305
Intended Use
To monitor the performance of the EasyStat and EasyBloodGas analyzers. Refer to Analyze QC in the Operator’s Manuals for detailed instructions. For in vitro diagnostic use.
Active Ingredients
EasyQC controls contain buffered bicarbonate and electrolyte solution equilibrated with precise mixtures of O2 and CO2. This control contains no preservatives or human-based materials.
Product Description
EasyQC is available for monitoring analyzer performance at different points in the clinical range. It is packaged in sealed glass ampules containing 1.7 mL of solution. Ampules are packaged thirty (30) per box.
Storage and Stability
Store control materials at 4–25°C. Changes in values may occur if stored at temperatures greater than 30°C for prolonged periods. DO NOT FREEZE. Discard any control material which exceeds the expiration date on the product label.
Recommended Use
Run quality control material once per shift before analyzing patient specimens. Run the controls at any time to verify analyzer performance. Use control material immediately after opening, following steps 1–5 in the instructions section of this sheet.
Expected Values:
Assigned values for each level of a specified lot of control material are based on the results of multiple tests run at Medica’s laboratories. Approximately 95% of all EasyStat and EasyBloodGas analyzers, operating according to specifications, are expected to fall within the rangeslisted on this sheet.